TIANCHI TC-210 veterinary diagnostic lab Manufacturer in TD

Publish Date 2019-08-22
TIANCHI TC-210 veterinary diagnostic lab Manufacturer in TD
Installation and disassembly
Connection between probe and main unit
Probe jack lies in the top of the right side, insert the probe into the jack correctly and turn it clockwise and wound it. There is only one probe jack which can be connected to all the supported optional probes.
1) Warning:
Avoid by all means unplugging or plugging the probe connector at state of log on in case the probe and main unit be damaged. Once the probe is connected with the main unit, do not unplug nor plug it at discretion in case poor contact happen.
2) Warning:
Must not touch the contact pin of the probe connector.
3) Warning:
The probe should be protected from felling off or crashing and the manufacturer assumes no responsibility for this kind of hazard.
4) Warning:
Please handle the device carefully.
Company:Henan Tianchi Instrument Equipment Co.,Ltd.
Tel:+86-371-56788280 +86-18903712477
Fax:+86-371-65715725
E-mail:tianchi-demi@ln2container.com
Website:www.ln2container.com
Twitter:TC18903712477
Facebook:18903712477
what's app:tianchiydg
QQ:648486900
TC-210
Model:TC-210
Probe:3.5MHz Mechanical sector
Displayed Depth (mm):60—190
Maximal detect depth(mm):≥140
Blind area(mm)≤8
Monitor size(inch):5.6 TFT-LCD
Display modes:B,B+B,B+M,M
Gray scales:256 levels
Image multiplying factor:*1.0,*1.2,*1.5,*2.0
Image storage:128-frame
Measuring functions:Distance,circumference,area,heart rate,gestation age
Battery capacity:2200mAh
Mainframe power:13W at n101on-charging operation/25W at charging operation
Adapter power consumption:45W
Character display:case history number,age
Faked color display:Gray,red,yellow,blue
Maintenance
1) Main unit maintenance
(1)Instrumentation environment should be according to "2.1 operation environmental requirements".
(2)If device enclosure needs cleaning, shutdown the device first and then wipe with alcohol sponges.
(3)Device should not be turn on and off frequently.
(4)When the device does not work for a long time, pack the device according to the instructions on the packing. Store it properly in the warehouse. The storage environment should accord to transportation and storage environmental requirements.
2) Probe maintenance
(1) Probe is an expensive part but frangible. Never hit it or drop it on floor. Please use medical ultrasound Gel during diagnoses. Do daily inspection on the probe enclosure to see whether it is cracked and to protect it from the water inrush.
(2) Check the probe regularly to see whether it is filled with clean medical castor oil, air in it will influence the image quality. If air bubbles appear, turn the probe oil filler hole up and screw off the sealing screw, turn the probe slowly to drive the air bubbles to the screw hole, infuse a little bit of castor oil into the screw hole with syringe with a needle, vent the air bubbles and fasten the sealing screw and finally clean the oil stain on the probe surface.